OPERATION
2-11
whenever the unit is triggered. In the triggered mode the unit will make a single
measurement per trigger.
STORE and RECALL
[Store] and [Recall] allow nine complete instrument setups to be saved in non-volatile memory. All the test
conditions, including binning configuration and open and short circuit compensation are saved.
Store To store a setup, press the [Store] key, which will display the 'StorE' message in
the display. Press the number of the stored setting ([1] - [9]) and the [ENTER]
key. [Store] [0] [ENTER] will give an error since setup 0 is the factory default
setup.
Recall To recall a stored setting, press the [Recall] key, which will display 'rcL'. Press the
number of the stored setting ([1] - [9]) and the [ENTER] key. [Recall] 0 returns the
default setup, including null cal values. If a 'CAL Err 4' appears during self test, or
a 'rcL Err' appears when recalling a setup, the stored setup was lost and must be
re-entered.